
152
ENSI Erfahrungs- und Forschungsbericht 2011
3.6. Pt concentration on the
specimen surface
Table 3 shows the Pt concentration measured on
the specimens from Tests 1 and 2. Although no
Pt particles were discovered in the SEM analysis
(apart from the pre-oxidised specimens of Test 2),
a very low but clearly measurable amount of Pt
has been detected on the surface of all specimens
using LA-ICP-MS. The Pt concentration was found
to be homogenous over the 120 measurement
points across the front and back side of each spe-
cimen. However, the uncertainty level marked in
the table amounts up to 50 %, mainly caused by
the calibration curve recorded at higher concent-
rations.
In spite of the high level of uncertainty, a clear
trend in Pt concentration between Tests 1 and 2
and also between the non pre-oxidised and pre-
oxidised specimens can be seen. The specimens
of Test 1 have less Pt deposited on the surfaces
than the ones from Test 2. This could be due to
the higher Pt injection rate in Test 2 resulting in a
higher Pt concentration in the high-temperature
water (Table 2). This result is also confirmed by the
more pronounced reduction of the ECP during Pt
injection in Test 2 (see Section 3.4). In case of both
tests, the non pre-oxidised specimens revealed lo-
wer Pt concentrations than the pre-oxidised. This
might be explained by the different size distribu-
tion of the oxide crystals, such as bimodal struc-
tures, leading to a higher surface roughness and
therefore more effective Pt deposition behaviour.
3.7. Summary
In a sophisticated high-temperature water loop
facility at PSI two experiments were performed in-
vestigating the Pt deposition behaviour under si-
mulated BWR conditions at two different (extre-
mely low) Pt injection rates and with differently
pre-treated coupon specimens. Taken together all
observations from the ECP measurements, SEM
and LA-ICP-MS analyses, these first tests show that
a more effective Pt deposition, resulting in a lower
ECP and therefore better protection against SCC,
could be achieved using a higher Pt injection rate
(corresponding to a higher Pt concentration in the
high-temperature water). Pre-oxidation of the spe-
cimens seems to additionally increase the Pt con-
centration on the specimen surface. Due to the
preliminary character of these results no further
conclusions can be drawn at this stage.
A more detailed analysis of the surface/oxide film
structure (e.g., using TEM) is necessary to get a
more conclusive picture of the Pt distribution on
stainless steel. Experiments with higher amounts
of Pt (closer to plant OLNC application conditions)
and different injection rates are underway to con-
firm the results and to get better insights into the
Pt deposition behaviour and processes behind. An
example image from such a test is shown in the
following sub-chapter.
3.8. Example image of a specimen
with higher Pt particle density
A SEM micrograph of a specimen from the latest
experiment, where a higher amount of Pt has been
injected into the high-temperature water loop
(665 µg) is shown in Figure 4. In contrast to the
first two tests (see above) a rather high density of
small Pt particles with a size distribution ranging
from 12 to 20 nm (15 nm in average) can be ob-
served on the oxide crystals. The analytical investi-
gation of this test (together with other compara-
ble tests) is still ongoing and conclusive results are
expected soon.
4. National collaborations
The collaboration and technology transfer on the
national level takes place within the Swiss nuclear
Test 1
Pt concentration [ng/cm
2
]
Test 2
Pt concentration [ng/cm
2
]
Non pre-oxidised Pre-oxidised Non pre-oxidised Pre-oxidised
10 ± 3 17 ± 6 11 ± 5 19 ± 6
Table 3:
Pt concentration on the
surface of specimens from
Tests 1 and 2, determined
by LA-ICP-MS.
Comentarios a estos manuales